Binary high-pressure liquid phase system (gradient)

NegotiableUpdate on 02/26
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The LC-10T binary high-pressure gradient system consists of four parts: high-pressure constant flow pump, automatic sampler, column temperature box, and ultraviolet detector. It is recommended to use Survey software for the chromatographic data workstation. It adopts a digital communication system, ultra wide dynamic range, and can perform reverse control of liquid chromatography on a computer. It can be widely applied in many analytical fields such as medicine, food, chemical industry, and environment. Provide customers with a comprehensive range of choices
Product Details


2 high-pressure constant flow pumps


One UV detector


1 automatic sampler


1 column temperature box


1 solvent tray





High pressure constant current infusion pump:


Infusion methodMicro volume series double plunger

Maximum infusion pressure6000psi

Flow setting range0.001~9.999ml/min (adjust flow rate in 0.001ml/min steps)

Error in flow setting value≤0.5%

Flow stability error≤0.2%RSD

pressure pulsation1% (0-100%, gradient of water/acetone aqueous solution 2) has

Pump sealing performanceLess than 0.1MPa, flow rate 1mL/min, pressure drop less than 5MPa

Time program functionhave

sizeW300*H135*D435mm


UV Visible Variable Wavelength Detector:


wavelength range190nm~700nm

Wavelength indication error≤±1nm

Wavelength repeatability error≤±0.1nm

Dynamic noise≤±0.75×10-5AU(甲醇,1mL/min,254nm,20℃)

static noise≤± 0.5 × 10-5AU (empty pool, response time 1 second, 20 ℃)

Dynamic baseline drift≤±1×10-4AU/h(甲醇,1mL/min,254nm,20℃)

Static baseline drift≤ 0.5 × 10-4AU/h (empty pool, response time 1 second, 20 ℃)

linear range≥104

Minimum detectable concentration≤ 4 × 10-9g/mL (naphthalene/methanol solution)

Qualitative repeatabilityRSD ≤0.1%

Quantitative repeatabilityRSD ≤0.5%

spectral bandwidth6nm

Flow pool volume10μL

optical path10mm

Time program functionhave

sizeW300*H135*D435mm


Column temperature box:


Temperature control: room temperature drops by 20 ℃ (under certain conditions, 3 ℃ -100 ℃)

Display mode: Actual temperature: Four digit LED display

Set temperature: Four digit LED display

Maximum resolution: 0.1 degrees Set resolution: 0.1 degrees

Temperature control accuracy: 0.1 degrees

Temperature stability: 0.1 degrees

Number of columns that can be accommodated: 2

Power supply: 220VAC 50HZ

Dimensions: 410 * 140 * 300mm

Heating box size: 360 * 40 * 40mm