CHOTT Tiron Line KF volumetric moisture titrator
Product Name:TitroLine KFCapacity based Karl Fischer titration instrument for water content
The latest TitroLine KF series Karl Fischer moisture titrator is an ideal tool for measuring sample moisture due to its lightweight size and fast reaction time.
Its broad practicality can almost encompass various application fields, such as:
Pharmaceutical industry: If tablets contain too much water, they will stick to each other; If the moisture content of the tablet is too low, it may be prone to breakage.
Chemical industry: high-purity solvents, quality control of various chemical products, inspection of raw materials and finished products, etc.
Food industry: For example, measuring the moisture content of food can avoid the problem of microbial growth.
Petroleum industry: The moisture content of oil products determines the practicality and quality of oil products
TitroLine KF includes the complete set of experimental equipment you need, placed in the same package, easy to assemble, and can be used in just ten minutes. A super large LCD display screen with backlight illumination, convenient for your use through a conversational program setting mode.
TitroLine KF is suitable for moisture testing in liquid, solid, gas, and viscous samples.
Experimental modes: sample titration, pure water titration, liquid standard titration, tartaric acid dihydrate titration, drying furnace blank, solvent blank.
Parameter settings: pre titration, extraction time, drift termination, endpoint delay time, endpoint current, polarization voltage, longest/shortest titration time, etc.
Report formats: Simple, Standard, and GLP.
The printing parameters include: titrant dosage, moisture content, statistical results (mean, standard deviation, relative variance), initial sample weight, date and time, sample number, titrant, blank value, drift value, titration time, experimental mode, titration parameters, calculation formula, signature field, etc.
Measurement formula and result unit: There are two formulas to choose from. The unit can be switched to%, ppm, mg, mg/L, mg/pc (pc stands for piece), and mL, while the blank value is mL.
Can be connected to various measurement accessories to extend measurement functions:
1. Titration control software;
2. Drying oven, used to determine the moisture content of insoluble or insoluble samples or substances that react with Karl Fischer reagent;
3. Homogenizer, used for directly stirring and crushing samples in the sample pool;
4. Balance;
5. Thermostatic titration cup;
6. Computers, printers, etc
In some quality management systems, traceability analysis results are strictly required, especially in the pharmaceutical industry. Therefore, Schott can provide you with the following table (optional attachment): DQ (Design Qualification) IQ(Installation Qualification),OQ(Operational Qualification)
Technical parameters:
| Range of moisture content |
100ppm to 100% H O |
| Compliant with standards |
DIN12 650, Part 5, And EN ISO8655-3 |
| Titration bottle |
Duran glass burette 20mL, with UV resistant cover |
| Valve and pipeline materials |
Three way two-way electromagnetic drive control valve, anti-corrosion material FEP with UV resistance function |
| Precision of allocation |
Error of 0.05%, measured according to EN ISO8655-6 |
| monitor |
LCD, Equipped with backlight illumination |
| External electrode |
Two platinum electrodes, potential: ± 1900mV; current: ± 200 μ A |
| Time |
30 seconds to a few minutes |
| RS232C |
Two sets of connection ports, can be connected to external computers or electronic scales |
| External dimensions/weight |
310x265x205mm (HxWxD)/approximately 3.2kg |
|