Phone
18516154700
-
Address
Building 30, No. 1525 Minqiang Road, Songjiang District, Shanghai
Shanghai Baipu Industrial Development Co., Ltd
18516154700
Building 30, No. 1525 Minqiang Road, Songjiang District, Shanghai
Druck Platinum Resistance Temperature Probe Calibration Instrument from the United States
(DPI811、DPI812)
Product features:
lDPI811/DPI812 is a tool for measuring and simulating RTD sensors, and is an ideal tool for inspecting RTD probes, indicators, recorders, and controllers. It supports 8 standard RTDs and is a full range, high reliability, easy-to-use handheld calibrator launched by General Electric Sensing and Measurement. It provides users with a good solution for process parameter testing and calibration
lMeasurement/simulation of thermal resistance signal output, 2, 3, 4-wire automatic detection, temperature testing and maintenance, transmitter calibration, circuit setting and diagnosis, switch value testing, support for IDOS pressure module, maintenance and upkeep * tools
l8 types of RTD temperature sensors (D100, Ni100, Ni120, PT50, PT100, PT200, PT500, PT1000) can be measured and simulated to automatically detect two-wire, three wire, and four wire RTDs
lCompatible with pulse type temperature transmitter
lStep/ramp output, Z-large/Z-small/average value, reading lock and other functions
Technical indicators:
lOperating environment: Temperature range of -10 to 50 degrees Celsius; Humidity 0-90%, non condensing
lProtection level: IP54, Dustproof and splash proof design
lVibration: DEF STAN 66-31, 8.81 and 8.4CAT III
lECM: EN61326-1; 1997+A1; 1998+A2; two thousand and one
lSecurity: EN61010; 2001, SEP grade, CE mark
lDisplay: 48mm × 48mm LCD with backlight, 99999 five bit resolution
lSize: 180 × 85 × 50 (mm)
lWeight: 500g
lBattery: 3 AA batteries, more than 50 hours for measurement, more than 10 hours for 24V circuit power supply
DPI 802/812/822Technical specifications for electrical signal interface:
lMilliampere measurement: 0 to 55mA, accuracy 0.02% rdg
l 24vCircuit power supply:+/- 10% (Z power supply 35mA)
lAutomatic switch value test: Capture open/close signals, 2mA test current
l HARTLoop resistance: menu selectable, default value 250
812The model also has the following functions:
lLoop calibration, providing RTD output and mA current measurement, suitable for calibration of various transmitter displays
l 24VSupply voltage to the transmitter and control circuit
lAutomatic temperature switch test, capturing on and off temperature values
l Hart resistorfunction
lAdditional pressure module can be connected, ranging from 70 mbar to 700 bar
DPI800series |
DPI811 |
DPI812 |
test type |
Platinum resistance probe |
Platinum resistance probe |
Indicator (pressure measurement) |
|
|
Calibration instrument (measurement/analog output) |
correct |
correct |
Temperature display instrument (dual input/differential) |
|
|
0-55MAMeasurement (providing 24V power supply) |
|
correct |
Switch quantity test |
|
correct |
HARTloop resistance |
|
correct |
IDOSIntelligent digital pressure module |
Note 1 |
Note 1 |
Programmable stepper/ramp output |
correct |
correct |
Reading retention/Z small/Z large/average/alarm/migration |
correct |
correct |
25Pressure unit, flow quantification, leakage testing |
Note 2 |
Note 2 |
1000Point data storage, RS232 interface |
Note 3 |
Note 3 |
Note 1: For the external pressure module, please refer to the IDOS list
Note 2: When used in conjunction with IDOS pressure module
Note 3: Refer to option E
IDOSPlug and play:
Each module can be interchanged between devices without the need for setup or instrument calibration
Measurement and simulation |
standard |
range |
accuracy |
Pt50(385) |
IEC 751 |
-200Up to 850 degrees |
0.5linear measure |
Pt100(385) |
IEC 751 |
-200Up to 850 degrees |
0.25linear measure |
Pt200(385) |
IEC 751 |
-200Up to 850 degrees |
0.6linear measure |
Pt500(385) |
IEC 751 |
-200Up to 850 degrees |
0.4linear measure |
Pt1000(385) |
IEC 751 |
-200Up to 400 degrees |
0.2linear measure |
D100(392) |
JIS1604-1989 |
-200Up to 650 degrees |
0.25linear measure |
Ni100 |
DIN43760 |
-60Up to 250 degrees |
0.2linear measure |
Ni120 |
MINCO7-120 |
-80Up to 260 degrees |
0.2linear measure |
Ohms |
|
0-4000 |
0.1-1.3Ω |
DPI812*Characteristics:
lMeasure 0-55.000mA with an accuracy of 0.02% of the reading plus 3 final words (including temperature effects of 10-30 degrees)
lTemperature coefficient: -10 to 10 degrees, 30 to 50 degrees, 0.002% PS/degree
lSwitch detection: open and close, 2mA current
lCircuit power output: 24V ± 10% (Z high current 35mA)
l HartLoop resistance: 250 Ω (menu selection)
lElectrical socket: 4mm socket
Item number |
product description |
DPI 811 |
Druck Platinum Resistance Temperature Calibration Instrument from the United States |
DPI 812 |
Druck Platinum Resistance Temperature Calibration Instrument from the United States |
Druck Platinum Resistance Temperature Probe Calibration Instrument from the United States