Welcome Customer !

Membership

Help

Beijing Mingnike Analytical Instrument Equipment Center
Custom manufacturer

Main Products:

smart-city-site>Products

Beijing Mingnike Analytical Instrument Equipment Center

  • E-mail

  • Phone

    13601000228

  • Address

    Room 1503, Building D, Xingyuan International, No. 222 Wangjing Xiyuan, Chaoyang District, Beijing

Contact Now

Flash evaporator

NegotiableUpdate on 04/23
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Flash evaporator features: The external sample transfer pipeline has adjustable heat tracing function to ensure that the sample will not condense, which can achieve high sample flux system blowing for liquid and gas samples and reduce cross contamination. High temperature carrier gas can be selected for system blowing
Product Details

S-200 flash evaporator

1、 Application:

This instrument is used for pre-treatment of liquid hydrocarbon substances before analysis, converting liquid samples into gaseous hydrocarbon substances with equal composition, constant temperature, constant pressure, and controllable flow rate for easy determination.

This product is suitable for sample processing before the following analysis:

1)Analysis of trace amounts of carbon monoxide and carbon dioxide in various industrial olefins, in conjunction with gas chromatography.

2)Liquid hydrocarbon composition analysis, used in conjunction with gas chromatography.

3)Analysis of other components.

IIFlash evaporatorparticular Point:

The external sample transfer pipeline has adjustable heat tracing function to ensure that the sample will not condense

Can achieve injection of liquid and gas samples

high throughput

System blowing reduces cross contamination

High temperature carrier gas can be selected for system purging

AComparison of similar domestic products:

Existing domestic products:

1)The flash evaporation device uses a chamber for heating and flash evaporation; The volume is too large, making it difficult to replace the sample, which will affect the next sample analysis;

2)The injection pipeline and flash chamber have not been effectively passivated; Affects the final analysis results of the sample;

3)The sample output pipeline of some models of products is not equipped with heat tracing output; Make the sample prone to condensation during transportation.


S-200Product:

S-200Overcoming the above problems, it automatically maintains a constant temperature for five minutes after connecting the power supply; Using foreign passivated pipelines and capillary injection, and entering the chromatographic injection analysis through a heat tracing pipeline. Thus overcoming the phenomenon of insufficient adsorption and vaporization of the sample in the pipeline.

S-200It is also equipped with a flash evaporation system for blowing and heating gas. After use, the carrier gas can be connected to the gas inlet and the carrier gas outlet can be connected to the liquid inlet for high-temperature blowing of the flash evaporation system, reducing cross contamination of residual samples.
B、 Technical features:

1)Liquid injection pressure: 100-400psi;

2)Design with two sets of user adjustable heaters. One heater controls the flash evaporation temperature, while another separate heater is used for insulation of the sample pipeline to prevent re condensation of the sample after gasification;

3)The maximum gasification temperature is 150 ℃;

4)Add high temperature protection to the external heat tracing pipeline to prevent overheating.

5)The external sample transmission pipeline has adjustable electric heating function to ensure that the sample will not condense;

6)Gas and liquidDual injection ports;

C、 Size:

wide*deep*tall=190*350*230