- Phone
-
Address
Room 706, Building 5, Xizi International Center, 898 Xiuwen Road, Minhang District, Shanghai (201199)
Shanghai Yifeng Instrument Equipment Co., Ltd
Room 706, Building 5, Xizi International Center, 898 Xiuwen Road, Minhang District, Shanghai (201199)
The EDM365 online environmental particulate/aerosol particle size spectrometer developed and produced by GRIMM Aerosol Technology Company in Germany uses laser scattering principle to simultaneously obtain the mass concentration values of PM10, PM2.5, and PM1.0 in the ambient atmosphere. Compliant with European standards EN12341 (PM10) and EN14907 (PM2.5) and certified by the US EPA (PM2.5). System integration design, equipped with meteorological parameter measurement sensors, can simultaneously monitor relative humidity, temperature, wind speed, wind direction, precipitation, and air pressure, and transmit monitoring data online. Optional gas sensor, capable of monitoring gaseous pollutants such as VOCs and NOx、O3And with CO.

Advantage
Fully automatic, anti oscillation, low maintenance
Outdoor movable stainless steel protective box with ventilation and heating equipment
3 PM particle sizes, TC value
Particle size analysis, PAH, and meteorological parameters (optional)
Can monitor relative humidity, temperature, and air pressure
GESYTEC protocol; Data recording card, RS-232 port
Filter membrane collection, can be used for chemical analysis and weight analysis (analysis)
No radioactive source, no SVC loss (Nafion dehumidification)

Application
Environmental monitoring station
Mobile PM monitoring
Analysis of aerosol particle size
Monitoring of hotspot areas
Identification of Pollution Sources
Parameter
Certification is the same as EDM180: EN12341, EN14907, US-EPA, GOSTR
Measurement principle: 90 ° laser scattering
Laser: Laser diode, wavelength 660nm
Power supply: Pmax.=60mW, normal 0.5/30mW CW (multi-channel)
Particle size channels: 11 nanochannels, 0.25/0.28/0.3/0.35/0.45/0.45/0.5/0.58/0.65/0.7/0.8/1.0,
20 micron channels, 1/1.3/1.6/2/2.5/3/3.5/4/5/6.5/7.5/8.5/10/12.5 15/17.5/20/25/30/32
Measurement mode:
EDM365-C PM10 and PM2.5 (Basic Version)
EDM365-CE PM10+TSP
EDM365-D PM10、PM2.5、PM1 和TC
EDM365-E 0.25~>32 μ m 31 particle size channels, or PM10, PM2.5, PM1, and TC
EDM365-MC 0.25~>32 μ m 31 particle size channels, as well as PM10, PM2.5, PM1, and TC
Particle concentration: 1~2 million p/L
Mass concentration: 0.1 μ g/m3~100 mg/m3
Weight calibration: extrapolation method, can be calibrated using filter membrane weighing method (EN12341, EN14907)
Sampling flow rate: 1.2LPM ± 5%
Cleaning gas flow rate: 0.3LPM, automatically running in standby mode
Dehumidification: Nafion drying tube, no heating
Repeatability: 5% maximum range
Operation: Panel operation or PC operation (software or super terminal)
LCD screen: 2 x 16 characters
Self check: automatically performed before each measurement
Measurement interval: 1 minute during continuous operation
Storage interval: 1min~1h
Data exchange: PC, RS232, data logger, data collector
Data output: LCD displays the average particle mass concentration per minute
Data interface: ASCII: RS-232 (baud rate 9600, bit 8, no parity check, stop bit 1 (Xon/Xoff);
GESYTEC: Baud rate 19200 1200, label and serial number optional
Data protocol: ASCII or GESYTEC
Memory card: 1MB and 4MB, battery support, storage interval is 1-60 minutes
Analog input: 3 (0-10V), resolution of 10 bits (ca. 10 mV)
Size: 70.0 × 48.9 × 27.1 cm
Weight: 36kg
Power supply: 220V-50Hz or 110V-60Hz
Power: max.P=150W, I (max)=1.4A
Maximum working altitude: 1000 meters, with PC control for flow regulation up to 5000 meters
Working environment: 4~40 ℃, RH<95%, no condensation
Storage and transportation conditions: -20~50 ℃, RH<95%, no condensation
Sample gas requirements: -20~60 ℃, RH<95%, no condensation
Pressure range: 0~-30 mbar (instantaneous)
Accessory
177B 32-bit software
365-SVC SVC update package, including dual sampling tubes for Nafion dehumidification and heating to 300 ℃ heater
365G various gas sensors (according to customer needs)
119 179 Special Small Parts
318 Second Year Spare Parts
157L humidity, temperature, and pressure sensor
158L humidity, temperature, pressure, wind direction and speed sensors
159L humidity, temperature, pressure, wind direction, wind speed, and precipitation sensors
Additional components for 180-DF 182 sampling tube (used in extremely cold areas)
160V2 PM10, PM2.5, and PM1 analog output converter, outputting 0~10V or 4~20mA, with a range of 0~6000 μ g/m3
The 1142.M5 data collector can transmit data to any PC to obtain and display real-time measurement data. The data from EDM365 and the connected sensors can be collected and combined into a set of data simultaneously, allowing for remote control of the instrument. 2GB of memory, upgradable to 4GB.
1142.A2 1MB 存储
1142. A4 4MB memory card
186 on-site testing components
Data recording system
The 1142.M5 data recording system integrates the data recording system and remote control system, connecting the GRIMM monitor with other sensors. The integrated interface is connected to the webpage and can be operated using TCP/IP protocol.
Automatic filter membrane sampling
For laboratory analysis, the sampling time and frequency can be set and changed by computer control, and more than 14 filter membranes (25-47mm) can be automatically replaced.
gas sensor
The gas system can be equipped with two different toxic gas sensors (CO, H)2S、SO2、NOxOr NH2)And a PID (photoionization detection) sensor can be installed in EDM365 and connected to the 1142.M5 data logger.
The 365G3xx gas sensor can continuously monitor various harmful organic and inorganic gases, oxygen, and flammable gases. Up to 5 sensors can be installed for emergency monitoring, industrial hygiene, and many other monitoring tasks. It is possible to combine standard gas sensors with special toxic gas sensors and PID for monitoring a wide range of toxic gases.


Gas sensor specifications:
| 365G313 | VOC/H2S/SO2Electrochemical sensor |
| 365G312 | VOC/H2S/CO electrochemical sensor |
| 365G315 | VOC/H2S/NO2Electrochemical sensor |
| 365G325 | VOC/CO/NO2Electrochemical sensor |
| 365G35C | VOC/CO2/NO2Electrochemical sensor |
Gas sensor specifications:
| monitoring parameters | Range/Resolution |
| VOCs | 0~199ppm/0.1ppm or 200~2000ppm/1ppm |
| CO | 10 ~ 2000ppm/10ppm |
| H2S | 0 ~ 100ppm/1ppm |
| SO2 | 0 ~ 20ppm/0.1ppm |
| NO2 | 0 ~ 20ppm/0.1ppm |
Software
The EDM series monitoring system runs using Labview software developed by GRIIM, which can display and calculate data in real time. All data is stored on memory cards or data loggers, and there will be no data loss. Due to the system being produced in Germany, all systems use the standardized GESYTEC network protocol. However, many other operating modes can be used through the GRIMM software package. Each device can be connected to a computer via RS232, and data logger can be used to collect data at intervals of 1 minute or more. Fully automated testing, control system monitors instrument performance and reports possible errors.