Pfeiffer Vacuum Gauge Basic Calibration System
In actual production, the measurement accuracy and repeatability of vacuum gauges directly affect the stability of the process and the quality of the product. Therefore, regular calibration of vacuum gauges is particularly important. In response to this application, Pfeiffer, a German company, has launched a vacuum gauge calibration system, which has been widely used in freeze-drying processes in the pharmaceutical industry
Pfeiffer Vacuum Gauge Basic Calibration System
Used for static and dynamic calibration
Calibration range 1.013-10-4 hPa
Easy to operate and flexible
Compliant with ISO/TS 3567 standard
Test up to 3 midsole vacuum gauges
Certified by DAkkS
Accept customer customization
Basic calibration system technical parameters:
|
model |
Basic calibration system |
Basic calibration system |
Basic calibration system |
||
|
accuracy |
According to different vacuum gauges |
||||
|
Cable length |
3 meters |
||||
|
Connect the vacuum gauge |
2x DN 16 ISO-KF |
3x DN 16 ISO-KF, |
6x DN 40 ISO-KF |
||
|
Gas flow rate |
5 X10-6-1E3 hPa·l/s |
||||
|
Air inlet |
DN 16 ISO-KF |
||||
|
measuring range |
1X10-4 – 1E3 hPa |
||||
|
Measurement range max |
1X103 hPa |
||||
|
Measurement range min |
1 X10-4 hPa |
||||
|
Operation method |
Manual |
||||
|
Electricity consumption |
230 VA |
||||
|
Front stage pump pumping speed 50 Hz |
0.9 m³/h |
||||
|
Nitrogen pumping speed |
67 l/s |
||||
|
extreme pressure |
< 1X10-7 hPa |
||||
|
weight |
27 kg |
||||
|
Data Download |
Chinese PDF |
Chinese PDF |
Chinese PDF |
||
Vacuum gauge calibration method:Based on the comparison of the measured values of the vacuum gauge to be calibrated and the standard vacuum gauge according to ISO/TS 3567 standard The vacuum gauge to be calibrated and the calibrated standard vacuum gauge are connected to a standardized vacuum chamber (according to ISO/TS 3567, specification b) and placed in the same vacuum environment
Provide static calibration and dynamic calibration, which can not only perform static calibration within a measurement range of>1 hPa, but also perform dynamic calibration within a measurement range of<1 hPa
Application of Pfeiffer Basic Calibration System
Freeze drying in the pharmaceutical and food industries
Coating system for optics, glasses, and decorative layers, diffusion barrier layer for PET bottles and packaging films, tool coating
Conduct process monitoring and control to determine the end of the process, in order to form stable quality and minimize scrap rates