Viscometers have multiple versions and are based on commonly used standards for rapid measurement of dynamic viscosity; It can also measure rheological properties for scientific research tasks involving further processing of liquid and viscosity data.Portable viscometerEasy to use, stable performance, and no need for thermometers or intermediate checks to check the viscosity of all on-site oils. It can be used to check the viscosity of various oils, from light spindle oil to heavy gear oil. Its flow cup body and outflow port are made of corrosion-resistant materials. Mainly used to determine the kinematic viscosity of Newtonian or approximate Newtonian fluid coatings, and can also be used for comparative measurements as needed.
The cup body of M384898 viscometer is made of high-quality copper material. The cup body of M384898 viscometer is made of synthetic rubberwood and has good corrosion resistance.
Installation method:
1. Place the flow cup in the circular ring of the cross arm.
2. The viscometer stand consists of an adjustable horizontal platform cross, with circular water bubbles attached to the cross arms. Adjust the horizontal bolts of the platform to center the water bubbles.
Instructions for use:
1. Cleaning treatment of flow cup: Dip a soft paper with good water absorption in a solvent that is easy to remove oil stains and repeatedly wipe the inner wall. Twist the paper into a rope shape and repeatedly pull it straight in the flow hole until it is clean.
2. Pour in the tested liquid: Block the outflow hole of the cup with your fingers (if there is corrosive liquid, use the stopper on the cup) and slowly pour the tested liquid into the cup until the liquid level protrudes from the upper edge of the cup. If there are bubbles, wait for the bubbles to float to the surface, and use a clean flat glass plate to push along the edge once to remove the protruding liquid level and bubbles, so that the horizontal plane of the tested liquid is on the same level as the upper edge of the flow cup.
3. Measure the outflow time: Release the blocked finger or push open the stopper, and start the stopwatch at the same time. The test solution flows out in a continuous line. When the line from the orifice starts to break, stop the stopwatch and record the stopwatch reading t.