Product Usage: The air-cooled xenon lamp aging test chamber uses a xenon arc lamp that can simulate the full sunlight spectrum to reproduce destructive light waves present in different environments. It can provide corresponding environmental simulation and light accelerated aging tests for scientific research, product development, and quality control.
| Equipment model |
Studio size (D * W * H) mm |
Dimensions (D * W * H) mm |
| LRHS-190F-SN |
500×760×500 |
1200×1270×1650 |
Technical Parameter
Sales Hotline: 021-608 99999
sexually ability finger treetop |
temperature range |
RT+10℃~70℃ |
| Humidity range |
65%~98%RH |
| Black panel temperature |
63℃~100℃ |
| temperature uniformity |
≤ 2.0 ℃ (in the dark) |
| temperature fluctuation |
± 0.5 ℃ (in the dark) |
| Humidity deviation |
+2、-3% RH |
| Spray nozzle aperture |
Ф0.8mm |
| Rainwater pressure |
0.12~0.15Mpa |
| Rainfall time |
1~999M, Continuous rainfall adjustable |
| Spray cycle |
18min/102min or 12min/48min (spray time/stop time) |
| photoperiod |
1~999h, m, s continuously adjustable |
| Sample tray size |
450×720mm |
| Distance between sample rack and lamp |
230~280mm |
| Wavelength |
290~800nm |
| radiation intensity |
≤550W/㎡ |
| Xenon lamp power |
2.5kW × 3 units=7.5kW |
| Power |
3.0kW |
| Humidification power |
2.0kW |
send use material material |
External material |
High quality A3 steel plate electrostatic spraying |
| Inner box material |
Imported high-grade stainless steel SUS304 |
| Temperature and humidity controller |
Taiwan Weilun touch screen+German Siemens PLC module |
| temperature and humidity sensor |
PT100 platinum resistance thermometer |
| Standard configuration |
3 light tubes, 1 layer of sample rack |
| Safety protection |
Compressor overpressure overload, fan motor overheating, equipment overheating, overall equipment under phase/reverse phase Equipment timing, leakage protection, water shortage protection, automatic shutdown after fault alarm, and other protections |
| supply voltage |
AC380V±10% 50Hz |
| Power |
14.5kW |
| Usage environment |
5℃~+28℃ ≤85% RH |
Note: 1. The above data were measured under good ventilation conditions at an ambient temperature of 25 ℃ 2. Customized air-cooled xenon lamp weather resistance test chambers can be customized according to specific user testing requirements 3. This technical information is subject to change without prior notice |