Chest drainage deviceDesigned and manufactured in accordance with the "YY/T 0583.2-2016 Disposable Thoracic Drainage Device Part 2: Dry Seal Valve" standard, suitable for testing the negative pressure status display, opening pressure, closing performance, and high negative pressure automatic release valve performance of the dry seal valve disposable thoracic drainage device.
Product Features:
·Equipped with a precise pressure testing system featuring absolute pressure sensors and vacuum generators, the testing accuracy has been greatly improved
·Embedded high-speed microcomputer chip control, simple and efficient human-computer interaction interface, providing users with a comfortable and smooth operating experience
·The 7-inch color touch screen, combined with intelligent data processing functions, provides users with a comfortable and smooth operating experience.
·A concise and intuitive user interface, convenient user-defined testing functions, and comprehensive data analysis and reporting capabilities.
·The operation process is simple and convenient, with a brand new UI interface design and intuitive user interface, which greatly improves the operating experience.
·The results are automatically judged and test data can be printed.
·Support quick viewing and printing of historical data.
·The instrument comes with a micro printer and powerful data storage function, capable of storing up to 50000 pieces of data
·The interface between Chinese and English can be freely switched
·Configure a standard RS232 communication port to facilitate external connection and data transmission between the system and the computer
·Energy saving and environmentally friendly design, beautiful appearance, ultra-low power supply meets national energy-saving and environmental protection requirements
Technical parameters of thoracic drainage device
Test range: -100 to 10 KPa (standard)
Accuracy: 0.5% FS
Gas source pressure: ≤ 0.7 MPa (provided by the gas source user)
Gas source interface: Φ 6mm polyurethane pipe
Dimensions: 390 mm (L) × 410 mm (W) × 190 mm (H)
reference standard
YY0583.2 中4.3.3、4.4.1、4.4.2、4.4.3、4.5.1、4.5.2
Product Configuration
Standard configuration: host, mini printer, air source connection
Note: The air source interface of this machine is made of 8mm polyurethane pipe; Gas source users should bring their own