The TA-200-H2 thermal conductivity hydrogen concentration analyzer adopts a 320 * 240 dot matrix LCD display and a standard 19 inch chassis, which can be integrated into a complete analysis system for convenience and speed. It is an industrial gas analyzer for hydrogen content in mixed gases developed using imported thermal conductivity sensors and new microcontroller technology. Widely used for rapid analysis of hydrogen content in various mixed gases such as petrochemicals, metallurgy, electricity, biological fermentation cracking, ammonia decomposition for hydrogen production, and electrolysis of water for hydrogen production.
Product Introduction:
TA-200-H2Thermal conductivity hydrogen concentration analyzerAdopting a 320 * 240 dot matrix LCD display and a standard 19 inch chassis, it can be integrated into a complete analysis system for convenient and fast useIndustrial gas analyzer for hydrogen content in mixed gases developed with imported thermal conductivity sensors and new single-chip technologyWidely used in petrochemicals, metallurgy, electricity, biological fermentation cracking, ammonia decomposition for hydrogen production, electrolysis of water for hydrogen production, etcRapid analysis of hydrogen content in various mixed gases.
TA-200-H2Thermal conductivity hydrogen concentration analyzerTechnical Specifications:
Measurement range:0-5%/10%/20%/50%/100% (customizable range)
Resolution: 0.01% or 0.1
Measurement accuracy: ≤± 2% F.S
Repeatability: ≤± 1% F.S
Stability: ≤± 1% F.S/7d
Response time: T90 ≤ 30s
Preheating time: ≤ 45 minutes
Sample gas flow rate: 400 ± 50mL/min
Sample gas pressure: 0.10Mpa ≤ inlet pressure ≤ 0.20Mpa
Sample gas temperature: 0-50℃
Environmental temperature: -5 ℃ to+50 ℃
Environmental humidity: ≤ 90% RH
Output signal: 4-20mA (can be customized according to requirements, must be specified when ordering)
Communication mode: RS232 or RS485 (communication protocol: Modbus RTU)
Working power supply: 100-240VAC 50/60Hz 1A
Dimensions: 3U Standard: 484 Width X 133 Height X 300 Depth 4U Standard: 484 Width X 177 Height X 380 Depth (mm)
Hole size: 3U standard: 484 width X 135 height 4U standard: 484 width X 179 height (mm)
Instrument weight: approximately 5kg
Measurement principle: Thermal conductivity sensor
Measurement medium: Conventional gases are acceptable (background gas composition should be provided when ordering)
Instrument lifespan: greater than five years (normal use)
Sensor lifespan: over two years (normal use)
Precision vacuum pump: used under negative pressure or normal pressure of the sample gas (must be specified when ordering)