Intelligent Battery Activation Instruments_TP_HVZH_ Technical Parameters
Date: 2025-12-29Read: 9

Yangzhou Topu Electric Technology Co., Ltd. provides the intelligent activation instrument for batteries with technical parameters of TPHVZH

The intelligent activation instrument for batteries has a very friendly human-machine interface. It can not only complete various settings and data queries under the prompts of the menu, but also store the discharge process data in the device's memory. Through the data interface, it can be transferred to a USB flash drive and analyzed by the dedicated software of the upper computer to generate the required curves and reports.

This instrument has complete protection functions, not only with sound and light alarms, but also with clear interface prompts.

This tester can serve as a discharge load when the battery is offline, and achieve constant current discharge of the set value by continuously regulating the discharge current. When discharging, the instrument will automatically stop discharging and record all valuable and continuous process real-time data when the terminal voltage of the battery pack or the individual voltage drops to the set lower limit value, or the set discharge time is reached, or the set discharge capacity is reached.

Product Features

Features

By using high-power winding resistors, the discharge current is more stable and fast.

There is a USB interface that can transfer data from the charging and discharging process to a USB flash drive and import it into a PC. PC data management software can analyze the process of battery discharge and generate corresponding data reports. Make data transfer more convenient.

Adopting intelligent microcontroller ARM control and 7-inch touch LCD display in both Chinese and English. The menu operation is simple and clear.

Automatic protection function, set the charging and discharging time to, charging and discharging capacity to, battery voltage below the set protection voltage, abnormal load connection, etc., automatically stop discharging and alarm, and automatically record the shutdown mode.

Multiple termination conditions for charging and discharging, including battery pack termination voltage, discharge capacity, and discharge time, ensure the safety of discharge testing.

Up to 10 automatic discharge charge cycles can be set

Due to the complexity of electrochemical reactions in batteries, as well as differences in materials, structures, manufacturing processes, and usage environments, there are significant differences in the characteristics of batteries produced by different manufacturers. Even batteries produced by the same manufacturer may have some variability in their individual characteristics. So far, there is no simple and effective method in the world that can quickly and accurately determine battery performance. The detection and failure prediction of battery performance remains a complex electrochemical measurement challenge.

Technical specifications

Applicable battery

lead-acid battery

Product model

A型

output voltage

1.5~16V

output current

2V battery 0~100A; 4/6V battery~60A; 12V battery 0~30A

working mode

Single machine use, constant current mode

protection performance

Input terminal overvoltage protection, LCD prompt;

Overcurrent protection, LCD prompt;

65 ℃ overheat protection, LCD prompt, buzzer alarm;

control accuracy

Discharge current ≤ ± 1%; Group terminal voltage ≤ ± 0.1%; Individual voltage: ≤± 0.05%

Data storage capacity

2Gbit FLASH

working environment

dissipate heat

Forced air cooling

temperature

Scope of work: -5~50 ℃ Storage temperature: -40~70 ℃

Humidity

Relative humidity 0-90% (40 ± 2 ℃)

Altitude

Rated altitude of 4000 meters

Noise

﹤60dB

working power supply

voltage

Single phase three wire system: 220V AC (-20%~+30%), frequency: 45-65Hz;

Withstand Voltage Test

Input - Chassis: 2200VDC 1min

Input output: 2200VDC 1min

Output - Chassis: 700Vdc 1min

Security

Meets EN610950

Wiring

communication input

National standard male socket, suitable for 1-1.5mm2 cables

DC output

25mm2 cable quick connect plug for instrument (red positive black negative)

Information organization: Yangzhou Topu Electric Technology Co., Ltd

TPXDC-220V-50A蓄电池恒流放电测试仪-1.jpg