E-mail
319059197@qq.com
-
Phone
15013668479
-
Address
East 3rd Floor, Building 15, Majialong Industrial Zone, Nantou Street, Nanshan District, Shenzhen
Shenzhen Yuanjian Sensing Technology Co., Ltd
319059197@qq.com
15013668479
East 3rd Floor, Building 15, Majialong Industrial Zone, Nantou Street, Nanshan District, Shenzhen
Product manual

product specification

product specification
|
Product Name: |
Online dust particle counter |
|
Product model: |
INS-DPC01 |
|
edition: |
V1.0 |
|
Developed by: |
He Genwen |
|
Reviewer: |
Yuan Chao |
The online particle counter can ensure uninterrupted operation 24 * 7 and has a long-term lifespan of ten years (note: under the premise of annual calibration). The online particle counter includes a laser sensor and an isokinetic sampling probe.
Online dust particle counter, used for detection in clean rooms and other clean environments. The counter is based on the principle of light scattering and applies photoelectric detection technology to measure and count the particle size and number of dust particles. The system has the advantages of good stability and accurate testing.
RS485 serial data transmission;
LCD touch screen display, high-end and atmospheric;
Small sensors monitor spatial micro particles;
Good system stability and accurate testing;
Has a long-term service life of ten years (note: subject to annual calibration and verification);
Provide 6 particle size channels: 0.3 μ m, 0.5 μ m, 1.0 μ m, 3.0 μ m, 5.0 μ m, and 10.0 μ m.

|
Pin |
name |
describe |
notes |
|
1 |
VCC |
Equipment power supply is positive |
DC24V |
|
2 |
GND |
Equipment ground |
—— |
|
3 |
NC |
—— |
—— |
|
4 |
485B |
Communication pins |
(RS485+) |
|
5 |
485A |
Communication pins |
(RS485- ) |
1. Parameter Table
|
Particle size range: |
0.3– 10 μm |
|
Channel size: |
Standard: 0.5, 5.0, μ m |
|
Counting efficiency: |
50% @ 0.5 μm; 100% for particles > 0.75 μm (per JIS) |
|
Current Speed: |
0.1 CFM (2.83 LPM) |
|
Laser source: |
Laser Diode , LASER Diode |
|
Zero count degree: |
<1 count / 5 minutes (per JIS) , <1 count/5 minutes (per JIS) |
|
Concentration control: |
5% at 2000000 particles per cubic foot |
|
Vacuum requirements: |
Built in vacuum pump |
|
Communication mode: |
RS-485 MODBUS protocol |
|
LED indicator light: |
Power supply, operation, flow rate, alarm |
|
Connector: |
MG 15EDGWB-3.81-5P |
|
Software: |
Pharmaceutical Industry Specific Version NETSCADA |
|
Shell: |
Stainless Steel , Stainless steel 304 |
|
Power Supply: |
DC 24V |
|
Size: |
13 cm (length) x 12 cm (width) x 8cm (height) |
|
Weight: |
0.5 kg |
|
power |
5W |
2. Environmental requirements:
Operation: Temperature range of 50 ° F to 104 ° F (10 ° C to 40 ° C)/humidity range of 20% to 95% without condensation. Storage: Temperature range of 14 ° F to 122 ° F (-10 ° C to 50 ° C)/humidity of 98% without condensation.
3. Accessories: power sampling probe, probe cap, pipeline, etc.
4. Calibration standards
1. Compliant with JJF1190-2008 Dust Particle Counter Calibration Specification:
2. Self purification time ≤ 10 minutes
3. Repeatability ≤ 10% FS
4. Particle size distribution error ≤± 30%
5. Particle concentration indication error ≤± 30% FS
6. Sampling flow error: 2.83L/min ≤± 5%
7. HLD780 long-life laser diode has a working time of ≥ 30000 hours.
1. Register description (the base address of this address is 1)
|
Modbus address |
Write |
Data type |
describe |
|
|
1 |
Read and write |
Integer |
The current address of the instrument (1-250, factory default 1), broadcast address 252 |
|
|
2 |
Read and write |
Integer |
Communication baud rate: 0-9600 (factory default), 1-19200, 2-38400 |
|
|
3 |
Read and write |
Integer |
Communication checksum: 0-no checksum (factory default), 1-even checksum, 2-odd checksum |
|
|
4 |
Read and write |
Integer |
Start stop instrument (set 1, start instrument; 0: stop instrument) or read the operating status of the instrument (1: start, 0: stop) |
|
|
5 |
Read and write |
Integer |
Counting delay (seconds) 0 100 seconds |
|
|
6 |
Read and write |
Integer |
Counting cycle (seconds) 2 |
|
|
7 |
read |
Integer |
standby |
|
|
8 |
Read and write |
Integer |
Running buzzer: Counting for one full cycle. Does the buzzer sound |
|
|
|
|
|
0-no sound (factory default), 1-sound |
|
9 |
Read and write |
Integer |
Alarm buzzer: Does the buzzer sound when the sampling period data alarm: 0-no sound (factory default), 1-sound |
|
10 |
read |
Integer |
Counter current count accumulation time (seconds) |
|
11-12 |
read |
Integer |
standby |
|
13 |
read |
Integer |
0.3um cumulative data is 16 bits lower |
|
14 |
read |
Integer |
0.3um cumulative data is 16 bits higher |
|
15 |
read |
Integer |
0.5um cumulative data low by 16 bits |
|
16 |
read |
Integer |
0.5um cumulative data is 16 bits higher |
|
17 |
read |
Integer |
Accumulated data of 1.0um is lower by 16 bits |
|
18 |
read |
Integer |
Accumulated data of 1.0um is 16 bits high |
|
19 |
read |
Integer |
3.0um cumulative data is 16 digits lower |
|
20 |
read |
Integer |
3.0um cumulative data is 16 digits higher |
|
21 |
read |
Integer |
Accumulated data of 5.0um is 16 digits lower |
|
22 |
read |
Integer |
5.0um cumulative data is 16 bits higher |
|
23 |
read |
Integer |
10.0um cumulative data is 16 digits lower |
|
24 |
read |
Integer |
10.0um cumulative data is 16 digits higher |
|
25 |
read |
Integer |
0.3um cycle data low 16 bits |
|
26 |
read |
Integer |
0.3um cycle data with 16 high bits |
|
27 |
read |
Integer |
Low 16 bits of 0.5um cycle data |
|
28 |
read |
Integer |
0.5um cycle data with 16 high bits |
|
29 |
read |
Integer |
Low 16 bits of 1.0um cycle data |
|
30 |
read |
Integer |
1.0um cycle data with 16 high bits |
|
31 |
read |
Integer |
3.0um cycle data low 16 bits |
|
32 |
read |
Integer |
3.0um cycle data with 16 high bits |
|
33 |
read |
Integer |
5.0um cycle data low 16 bits |
|
34 |
read |
Integer |
5.0um cycle data with 16 high bits |
|
35 |
read |
Integer |
10.0um cycle data low 16 bits |
|
36 |
read |
Integer |
10.0um cycle data with 16 high bits |
|
37 |
read |
Integer |
Current traffic data/100=current traffic |
|
38 |
read |
Integer |
Temperature (with 1 decimal place, value 100 represents 10.0 ℃) |
|
39 |
read |
Integer |
Humidity (with 1 decimal place, value 100 represents 10.0%) |
|
40 |
read |
Integer |
Atmospheric pressure (with 2 decimal places, a value of 10000 represents 100.00Kpa) |
|
41 |
Read and write |
Integer |
0.3um correction factor (%): 0~10000%, default 100% |
|
42 |
Read and write |
Integer |
0.5um correction factor (%): 0~10000%, default 100% |
|
43 |
Read and write |
Integer |
1.0um correction factor (%): 0~10000%, default 100% |
|
44 |
Read and write |
Integer |
3.0um correction factor (%): 0~10000%, default 100% |
|
45 |
Read and write |
Integer |
5.0um correction factor (%): 0~10000%, default 100% |
|
46 |
Read and write |
Integer |
10.0um correction factor (%): 0~10000%, default 100% |
|
47 |
Read and write |
Integer |
Temperature correction value (℃): -100~100 (-10.0~10.0 ℃) |
|
48 |
Read and write |
Integer |
Humidity correction value (%): -100~100 (-10.0~10.0%) |
|
49 |
Read and write |
Integer |
Atmospheric pressure correction value (kPa):-1000~1000(-10.00~ 10.00kPa) |
|
50 |
read |
Integer |
standby |
|
51 |
Read and write |
Integer |
0.3um alarm enabled 0: No alarm (factory default), 1: Alarm |
|
52 |
read |
Integer |
0.3um alarm status 0: No alarm, 1: Alarm |
|
53 |
Read and write |
Integer |
0.3um alarm value low by 16 digits |
|
54 |
Read and write |
Integer |
0.3um alarm value high by 16 digits |
|
55 |
Read and write |
Integer |
0.5um alarm enable 0: No alarm (factory default), 1: Alarm |
|
56 |
read |
Integer |
0.5um alarm status 0: No alarm, 1: Alarm |
|
57 |
Read and write |
Integer |
0.5um alarm value low by 16 digits |
|
58 |
Read and write |
Integer |
0.5um alarm value high by 16 digits |
|
59 |
Read and write |
Integer |
1.0um alarm enabled 0: No alarm (factory default), 1: Alarm |
|
60 |
read |
Integer |
1.0um alarm status 0: No alarm, 1: Alarm |
|
61 |
Read and write |
Integer |
1.0um alarm value low by 16 digits |
|
62 |
Read and write |
Integer |
1.0um alarm value high by 16 digits |
|
63 |
Read and write |
Integer |
3.0um alarm enable 0: No alarm (factory default), 1: Alarm |
|
64 |
read |
Integer |
3.0um alarm status 0: No alarm, 1: Alarm |
|
65 |
Read and write |
Integer |
3.0um alarm value low by 16 digits |
|
66 |
Read and write |
Integer |
3.0um alarm value high by 16 digits |
|
67 |
Read and write |
Integer |
5.0um alarm enable 0: No alarm (factory default), 1: Alarm |
|
68 |
read |
Integer |
5.0um alarm status 0: No alarm, 1: Alarm |
|
69 |
Read and write |
Integer |
5.0um alarm value low by 16 digits |
|
70 |
Read and write |
Integer |
5.0um alarm value high by 16 digits |
|
71 |
Read and write |
Integer |
10.0um alarm enable |
|
|
|
|
0: No alarm (factory default), 1: Alarm |
|
72 |
read |
Integer |
10.0um alarm status 0: No alarm, 1: Alarm |
|
73 |
Read and write |
Integer |
10.0um alarm value low by 16 digits |
|
74 |
Read and write |
Integer |
10.0um alarm value high by 16 digits |
|
75 |
Read and write |
Integer |
Temperature alarm enabled 0: No alarm (factory default), 1: Alarm |
|
76 |
read |
Integer |
Temperature alarm status 0: No alarm, 1: Alarm |
|
77 |
Read and write |
Integer |
Temperature alarm value low by 16 digits |
|
78 |
Read and write |
Integer |
Temperature alarm value is 16 digits high |
|
79 |
Read and write |
Integer |
Humidity alarm enabled 0: No alarm (factory default), 1: Alarm |
|
80 |
read |
Integer |
Temperature alarm status 0: No alarm, 1: Alarm |
|
81 |
Read and write |
Integer |
Humidity alarm value low by 16 digits |
|
82 |
Read and write |
Integer |
Humidity alarm value is 16 digits high |
In addition: Modbus data bit 8, stop bit 1
Special note: Address 252 is used for broadcasting addresses, as long as it is used for the following situations
(1) The situation where there is only one instrument on the entire bus
At this point, the communication address of the current instrument can be set
(2) Start or stop all instruments on the entire bus
If there are multiple instruments on the bus, setting the address (register 0) with 252 address on the bus will cause all instruments on the bus to have the same address, and future communication will not be able to correctly identify a certain instrument, resulting in a malfunction of the entire system.
2. MODBUS commands supported by instruments
|
function code |
role |
describe |
|
03 |
Read and hold register (16 bit register) |
Read one or more hold registers |
|
06 |
Write a single 16 bit register |
Set the current value of a hold register |
|
16 |
Write multiple 16 bit registers |
Set one or more hold registers |
1. The product is equipped with a laser emitter inside, and unauthorized maintenance may result in accidental exposure of operators to laser radiation. It should be repaired by specialized personnel from the manufacturer;
2. Prohibit opening the product casing for use;
3. The product is not in direct contact with the monitoring environment (such as installed in the equipment chassis). In order to obtain accurate sampling and measurement results, please connect the air inlet to the external probe of the chassis, and the length of the connecting hose between the external probe and the inlet should be controlled within 30cm;