CSI-Z506 radial support force detector (bath constant temperature type), (nickel titanium shape memory alloy self expanding type)Radial support force tester for vascular stent
Radial support force tester for vascular stent
1、 Instrument Introduction:
The water bath type radial support force tester can keep the waterproof pressure grip head in a water bath at 37 ℃ (or other required temperature) for a long time, meeting the requirements for constant temperature radial force testing of temperature sensitive metals or other materials, especially for radial force testing of stent medical device products such as nickel titanium alloy and shape memory alloy.
2、 Key parameters
Parameter items |
Specifications |
control system |
PLC+Windows control system |
operating interface |
12 inch embedded industrial computer, switching between Chinese and English |
test software |
Monitoring and parameter settings, historical data, experimental reports, curve monitoring, software upgrades |
printer |
WiFi connection to printer |
force sensor |
0~100N accuracy: ± 0.5% FS (optional ± 0.1% FS) |
Test sample length |
0~100mm (other length specifications can be customized) |
speed range |
0~5mm/s (adjustable) |
test method |
Iris type rolling radial loading device |
constant temperature water bath |
RT~50 ℃ Temperature Control Accuracy: ± 2 ℃ |
power supply |
AC 220V/50Hz |
Overall dimensions |
680*750*1100mm |
3、 Execution standards and applicable scenarios
Compliant with standards:
T/CSBM 0018-2021 "Test Method for Shape Recovery Ability of Nickel Titanium Shape Memory Alloy Self Expanding Vascular Stent"
YY/T1660-2019 "Radial Load Test Method for Balloon Dilation and Self Dilation Vascular Stents"
Typical applications:
Batch quality inspection of production line;
Third party laboratory certification.

4、 Example of Operation Process
1) If the vascular stent has been loaded, immerse the vascular stent system in a constant temperature water bath at 37C ± 2C for no less than 2 minutes, and release the vascular stent from the sheath in the constant temperature water bath; If the bracket is not loaded, this operation is not required.
2) For self expanding variable diameter vascular stents, it is necessary to evaluate the radial support force of different segments.
3) The temperature of the testing device is maintained at 37C ± 2C; the initial aperture is larger than the natural diameter of the testing bracket.
4)Place the released vascular stent in the testing hole of the device at a constant temperature for at least 2 minutes.
5) According to the status of the vascular stent, slowly reduce the aperture of the testing device as much as possible to the preloaded diameter of the vascular stent or reduce the diameter by at least 50%, and then slowly increase it to the unloaded state (recommended moving speed: 0.1mm/s~0.5mm/s).
6) Measure the force value of expansion or compression corresponding to the expected size of the implanted blood vessel diameter during the testing process, and observe whether the vascular stent can restore its initial shape after the external force is removed.
5、 Configuration List
1 host;
1 set of testing software;
1 copy of instruction manual;
1 certificate of conformity;
1 warranty card;
One copy of the delivery receipt;
1 nameplate;
1 power cord;
1 set of hexagonal wrench;
Several brochures;